How to fill a cartridge
-
Gather your supplies
- Room temperature insulin
- Cartridge set
- Alcohol wipe
-
Prepare Cartridge
Complete steps before attaching vial.
- Pull fill rod down.
- Push fill rod back up completely.
- Pull fill rod down to desired volume.
-
Attach Insulin Vial
- Place insulin vial on a flat surface.
- Push vial adapter down onto vial.
-
Fill Cartridge
- Push fill rod down to force all air into vial and maintain pressure.
- Turn set upside down.
- Slowly pull fill rod back to desired volume.
-
Remove Insulin Vial
- Pull vial out from vial adapter.
-
Inspect for Air
Air takes space where insulin should be and can affect insulin delivery.
- Check all sides of cartridge for air.
-
Remove Air
- Hold upright and tap so air rises to top.
- Slowly push air up and out of cartridge.
- Repeat as needed.
-
Remove Cartridge from Set
- Unscrew fill rod.
- Press tab to release cartridge.
- Unscrew vial adapter.